Evaluating Oxygenation, Blood Gas Analyzers & QA/QC
Objective 1.3.1 — Distinguish hypoxemia from hypoxia and stage PaO₂ severity (with age adjustment); calculate CaO₂ and the A–a gradient using the alveolar air equation; describe the SANZ, Severinghaus, and Clark electrodes and lab procedures for ABG analysis; explain co-oximetry and the four common hemoglobin species; apply point-of-care testing (POCT / i-STAT); and follow regulatory, calibration, and quality-control protocols (CLIA · CAP · Levey-Jennings · Westgard rules).
Evaluating Oxygenation
| Term | Definition |
|---|---|
| Hypoxemia | Abnormally low level of O₂ in the blood |
| Hypoxia | Deficient amount of O₂ reaching the tissues — global or local |
Assess for Hypoxemia — PaO₂ Ranges
Normal PaO₂ ranges for patients less than 60 years old, breathing 21% O₂ at sea level.
| PaO₂ (mmHg) | Interpretation |
|---|---|
| 80 – 100 | Normal |
| 60 – 79 | Mild hypoxemia |
| 45 – 59 | Moderate hypoxemia |
| < 45 | Severe hypoxemia |
Age Adjustment — Patients Over 60
Rule of thumb: for patients over 60, subtract additional years from 80.
Example: 66-year-old patient → 80 − 6 = 74 mmHg PaO₂ is the expected normal.
Assess for Hypoxia
Hypoxia is evaluated using two main calculations:
| Calculation | Abbreviation | Tells You |
|---|---|---|
| Arterial oxygen content | CaO₂ | Total O₂ delivered to the tissues |
| Arterial O₂ tension gradient | P(A–a)O₂ | Difference between alveolar and arterial PO₂ |
Arterial Oxygen Content (CaO₂)
The best measurement of oxygen delivered to the tissues — and the best index of oxygen transport. CaO₂ estimates O₂ carried by hemoglobin plus O₂ dissolved in plasma.
CaO₂ = (Hb × 1.34 × SaO₂) + (PaO₂ × 0.003)
First term = O₂ in the RBC · Second term = O₂ in plasma
| Component | Meaning |
|---|---|
| Hb × 1.34 | Each gram of Hb carries 1.34 mL O₂ when fully saturated |
| × SaO₂ | Multiply by saturation (decimal) |
| PaO₂ × 0.003 | Dissolved O₂ in plasma (Henry's law constant) |
Normal value: 17 – 20 vol % (mL/dL)
Worked Example
Calculate the CaO₂ for a patient with: SaO₂ 98%, PaO₂ 90 torr, Hb 14 g/dL.
CaO₂ = (14 × 1.34 × 0.98) + (90 × 0.003)
= (18.388) + (0.27)
≈ 18.66 vol % (mL/dL)
Within the normal range of 17–20 vol %.
A–a Gradient (A–aDO₂, P(A–a)O₂)
The A–a gradient measures the difference (gradient) between alveolar and arterial PO₂.
Best done after the patient has been on 100% O₂ for 20 minutes or more.
A–aDO₂ = PAO₂ − PaO₂
| A–a Gradient on 100% O₂ | Interpretation |
|---|---|
| 25 – 65 torr | Normal |
| 66 – 300 torr | V/Q mismatch |
| > 300 torr | Shunting |
Alveolar Air Equation — PAO₂
Calculates the partial pressure of O₂ in the alveoli.
PAO₂ = (P_B − P_H₂O) × FiO₂ − (PaCO₂ ÷ 0.8)
| Constant | Normal Value |
|---|---|
| P_H₂O (water vapor pressure) | 47 torr |
| R (respiratory exchange ratio) | 0.8 |
Normal value varies directly with the patient's FiO₂ and barometric pressure (P_B).
Worked Example
Calculate the PAO₂ for a patient on 50% O₂ with PaCO₂ 40 torr and P_B 747 torr.
PAO₂ = (747 − 47) × 0.50 − (40 ÷ 0.8)
= (700)(0.50) − 50
= 350 − 50
= 300 torr
A–a Gradient Worked Example
What would the A–a gradient be for the patient above if the PaO₂ was 80 torr?
A–aDO₂ = PAO₂ − PaO₂
= 300 − 80
= 220 torr → V/Q mismatch range
Blood Gas Analyzers — Electrode Functional Operation
Electrode Terminology
| Term | Meaning |
|---|---|
| Anode | Positive pole (Ag/AgCl) |
| Cathode | Negative pole (Platinum) |
| Oxidation | A loss of electrons by a particle |
| Reduction | A gain of electrons by a particle |
The Three ABG Electrodes
| Electrode | Measures |
|---|---|
| SANZ electrode | pH |
| Severinghaus electrode | PCO₂ |
| Clark electrode | PO₂ |
Mnemonic: "Some Senior Clinicians" — SANZ → pH, Severinghaus → PCO₂, Clark → PO₂.
Laboratory Procedures for ABG Analysis
- Obtain the sample.
- Ensure all air bubbles are removed from the syringe.
- Mix the sample thoroughly.
- Introduce sample into the analyzer.
- As the sample runs through the analyzer, watch for air bubbles or clots.
- Print, record, and report results in accordance with local policy.
- Check results for errors:
- If inaccurate → run sample through another analyzer to compare results.
- If another analyzer is not available → place sample on ice.
- Run a control through the analyzer to check for accuracy.
- Temperature correction — results CAN be corrected if patient temperature is other than 37 °C (98.6 °F).
Co-oximeter
Spectrophotometry — How It Works
| Root | Meaning |
|---|---|
| Spectro- | spectrum |
| -photo- | light |
| -metry | measurement |
Light is transmitted at specific wavelengths through the blood and received by a photodetector on the opposite side.
Light Transmission Behavior
When light hits blood it can be:
- Absorbed (optical density)
- Pass through (inverse to light absorbed)
- Reflected (backscatter oximetry)
Lambert-Beer Law
The Lambert-Beer law determines hemoglobin saturation and concentration.
Each form of hemoglobin has its own unique absorption / transmission spectrum (isobestic points).
Isobestic points — used to determine hemoglobin concentration regardless of its saturation.
Co-Oximetry — 8 Wavelengths, Multiple Hemoglobin Species
Co-oximetry uses eight light wavelengths to differentiate between various hemoglobin species:
| Species | Description |
|---|---|
| Hb | Deoxygenated |
| HbO₂ | Oxygenated |
| MetHb | Methemoglobin (Fe³⁺) |
| HbCO | Carboxyhemoglobin |
| SHb | Sulfhemoglobin |
SaO₂, Total Hemoglobin, and Functional vs. Fractional
SaO₂ is the percentage of hemoglobin bound with oxygen.
Total Hemoglobin (tHb, g/dL) is comprised of functional and fractional SaO₂:
| Component | Meaning |
|---|---|
| Functional SaO₂ | The SpO₂ equivalent — Hb bound to oxygen, typically taken from a pulse-ox reading |
| Fractional SaO₂ | The deoxygenated Hb species (all other forms of Hb) |
Normal Total Hemoglobin Values
| Patient | Normal tHb (g/dL) |
|---|---|
| Adult male | 13.5 – 18.0 |
| Adult female | 12.0 – 16.0 |
| Newborn | 14.0 – 24.0 |
Carboxyhemoglobin (HbCO) and Methemoglobin (MetHb)
Carboxyhemoglobin — HbCO
- Caused by smoke inhalation / CO exposure
- Normal range: 1.5 – 9.0 % of tHb
Methemoglobin — MetHb
- Normal: 1 % or less
- Methemoglobinemia = abnormal blood state where MetHb is > 1%
| Cause | Setting |
|---|---|
| iNO (inhaled nitric oxide) | ICU |
| Hurricane topical anesthetic | Bronchoscopy |
| Nitroglycerin | — |
Treatment of methemoglobinemia: METHYLENE BLUE.
Point-of-Care Testing (POCT)
Definition: Laboratory testing performed outside the central laboratory (at the patient's bedside).
Classified as "Moderately Complex" according to CLIA.
Benefits of POCT
- Shortened turnaround time
- More timely intervention
- Correlates results with patient status at the bedside
- Reduced errors related to handling, labeling, and reporting
- Small sample size — only 2–3 drops of blood
POCT Equipment — Handheld Analyzer
- Cartridge port
- Numerical keypad
- 50-patient sample storage
- IR transmitter
- Battery powered
- Printer
- Calibration card
- Cartridges
i-STAT Procedure
Obtaining the Sample
- Collect blood specimen.
- Syringe with anticoagulant → test within 5 minutes.
- Capillary tubes → test within 3 minutes.
- Gently mix sample in syringes or collection tubes.
- Practice sample-handling precautions per organizational policy.
Analyzing the Specimen
- Fill the cartridge via the sample well using slow, steady pressure — 2–3 drops.
- Close the cartridge, press on the tab until it snaps.
Quality Assurance (QA)
Quality assurance — a systematic process designed to:
- Monitor
- Document
- Regulate the accuracy and reliability of laboratory measurements
Regulatory and Accreditation Agencies
| Acronym | Agency |
|---|---|
| CLIA | Clinical Laboratory Improvement Act / Amendments |
| CAP | College of American Pathologists |
| TJC | The Joint Commission |
| CLIP | Clinical Laboratory Improvement Program — military version of CLIA |
| CMS | Centers for Medicare & Medicaid Services |
| NCCLS | National Committee of Clinical Laboratory Standards |
| Local OIs | Local operating instructions |
Preventive Maintenance
- Clean blood spills with appropriate cleaning solution to eliminate contamination hazard
- Replace waste bottles
- Check supply levels and replace if necessary:
- Wash and cleaning solutions
- Calibration buffer solutions
- Calibration gas tank pressure (CO₂)
Always wear gloves and eye protection.
Non-Scheduled Maintenance — Replace as Required
- Electrodes
- Electrode solution
- Electrode membrane
- Pump tubing
- Clean dust filters — replace as needed
Calibration
Calibration — equipment is adjusted/corrected to match the control standards. Performed on electrodes before analyzing samples to establish the accuracy of results.
Calibration standards = specific blood gas values used to set the machine to read linearly over the expected range of results.
Calibration Levels
| Level | What It Does | When |
|---|---|---|
| 1-point | Calibrates pH, pCO₂, pO₂ to one point | Before every blood sample is run · automatically every 30 minutes · manually after maintenance |
| 2-point | Calibrates electrodes at two points (low and high) | Controls every 8 hours · manually after maintenance |
| 3-point | Checks the machine for linearity | Every 6 months |
Quality Control (QC)
Quality Control — a test performed to determine the accuracy and precision of a device against a known standard.
| Term | Meaning |
|---|---|
| Accuracy | How closely the measured results reflect the actual value |
| Precision | An index of dispersion of repeated measurements |
Internal Quality Control (Analyzer Controls)
- Controls = samples with known values run to ensure the analyzer is operating correctly.
- Assures the reliability of blood-sample values.
- Values should be within the manufacturer's standard deviation.
| Setting | Frequency |
|---|---|
| 24/7 lab (respiratory therapy) | Once every 8 hours |
| Pulmonary lab (8-hour shift) | Once before the start of the normal duty day |
Control Statistics
Control results are maintained in statistical databases and evaluated to detect changes in analyzer performance.
Levey-Jennings Charts & Westgard Rules
Levey-Jennings Chart
A graphic representation of each control run on each electrode, plotted around the mean with ±1, ±2, ±3 standard-deviation lines.
Westgard Rules
Specific criteria and actions to be taken based on the results of quality control samples.
| Result Status | Action |
|---|---|
| In control | No action required |
| Random error | Monitor closely |
| Out of control | Must perform corrective action |
Random Error
- One measurement outside of 2 SD → monitor closely.
"Out of Control" — Perform Corrective Action
| Westgard Rule | Trigger |
|---|---|
| Rule of 2's | Two consecutive measurements outside 2 SD |
| Rule of 3 | One measurement outside 3 SD |
| Rule of four 1's | Four consecutive measurements outside 1 SD |
| Rule of 10's | 10 consecutive control measurements fall on one side of the mean |
External Quality Control (Proficiency Testing)
A system that compares the accuracy of results from a lab with results obtained from other labs.
- Identical blind (unknown) samples are sent from an outside agency (CAP) to participating labs.
- Samples are analyzed and returned to the sending agency.
- Results are compared to the known values and to results from other labs.
College of American Pathologists (CAP)
Outside agency introduced in the late 1940s.
CAP's Purpose
- Perform impartial and fair inspections
- Assess the state of the art in all clinical laboratory practice
- Promote laboratory improvement through peer comparison
- Satisfy regulatory requirements of credentialing agencies
- Grant accreditation
CAP Accreditation Requirements
- Quality management program
- Laboratory safety plan
- Document control plan
- Competency assessment program
- Laboratory director oversight documentation
- Specifies requirements for a lab information system
- Standardizes the operation of all laboratories
CAP inspections are every TWO YEARS.
CAP External Quality Control
Systematically compares accuracy of results obtained from multiple laboratories:
- Identical blind samples sent from CAP.
- Consists of 5 – 10 control samples.
- Limits and standard deviations are only known to CAP.
- Samples analyzed and returned to CAP.
- Results compared to known values and to other labs.
Quick Reference
Hypoxemia Stages (under 60 y/o, RA, sea level)
80–100 Normal · 60–79 Mild · 45–59 Moderate · <45 Severe
Age-Adjusted PaO₂
Over 60 → 80 minus (years over 60). A 70-year-old: 80 − 10 = 70 mmHg expected.
CaO₂ Formula
CaO₂ = (Hb × 1.34 × SaO₂) + (PaO₂ × 0.003) — normal 17–20 vol %.
Alveolar Air Equation
PAO₂ = (P_B − 47) × FiO₂ − (PaCO₂ / 0.8).
A–a Gradient on 100% O₂
25–65 normal · 66–300 V/Q mismatch · >300 shunt.
Three Electrodes
SANZ → pH · Severinghaus → PCO₂ · Clark → PO₂.
Anode vs. Cathode
Anode = positive (Ag/AgCl) · Cathode = negative (platinum). Oxidation = LOSE electrons · Reduction = GAIN electrons (OIL RIG).
Hb Species (Co-oximetry)
Hb · HbO₂ · MetHb (Fe³⁺) · HbCO · SHb — differentiated by 8 wavelengths via the Lambert-Beer law.
Hb-Carrying Capacity
1.34 mL O₂ per gram of Hb.
Normal tHb
Male 13.5–18 · Female 12–16 · Newborn 14–24 g/dL.
HbCO and MetHb
HbCO normal 1.5–9.0% of tHb (smoke / CO). MetHb normal ≤1%; >1% = methemoglobinemia. Causes: iNO, Hurricane spray, nitroglycerin. Treatment: methylene blue.
POCT
Outside the central lab · CLIA "moderately complex" · 2–3 drops · syringe sample tested <5 min, capillary <3 min.
Calibration Cadence
| Level | Frequency |
|---|---|
| 1-point | Before every sample · auto every 30 min |
| 2-point | Every 8 hours |
| 3-point | Every 6 months (linearity) |
Accuracy vs. Precision
Accuracy = closeness to true value · Precision = dispersion of repeated values.
Westgard Cheat Sheet
| Rule | Trigger | Action |
|---|---|---|
| 1 measurement >2 SD | Random error | Monitor closely |
| 2 consecutive >2 SD | Rule of 2's | Corrective action |
| 1 measurement >3 SD | Rule of 3 | Corrective action |
| 4 consecutive >1 SD | Rule of four 1's | Corrective action |
| 10 consecutive on one side of mean | Rule of 10's | Corrective action |
CAP At-A-Glance
Established late 1940s · grants accreditation · inspections every 2 years · external proficiency = 5–10 blind samples sent to labs.
Regulatory Alphabet Soup
CLIA · CAP · TJC · CLIP (military CLIA) · CMS · NCCLS.