Oxygen Delivery Devices

O2 indications, contraindications, and hazards; low- and high-flow delivery devices; Rule of Fours; Venturi entrainment; HFNC; and oxygen analyzer calibration.

Listen: Oxygen Delivery Devices

0:00
--:--

Oxygen Delivery Devices

Objectives - Describe indications for oxygen therapy; administer oxygen utilizing low-flow and high-flow therapy devices; describe flow rate and FiO2 range for each oxygen device; perform a two-point calibration on an oxygen analyzer.


Indications for Oxygen

Documented Hypoxemia

Confirmed by ABG:

  • PaO2 less than 60 mmHg or SaO2 less than 90% on room air (21% FiO2)
  • PaO2 or SaO2 below desirable range for specific clinical situations, such as pneumonia

Acute Care Situations in Which Hypoxemia Is Suspected

Examples include:

  • Cor pulmonale
  • Congestive heart failure
  • Severe trauma (severe blood loss)
  • Acute myocardial infarction

Short-Term Therapy or Surgical Intervention

  • Post-anesthesia recovery

Contraindications for Oxygen

With few exceptions, no specific contraindication to O2 therapy exists when indications are present.

Certain delivery devices are contraindicated:

  • Nasal cannulas and nasopharyngeal catheters in pediatric and neonatal patients with obstruction

Hazards of Oxygen

  • PaO2 greater than or equal to 60 mmHg: ventilatory depression may occur rarely in spontaneously breathing patients with elevated PaCO2 (COPD)
  • With FiO2 greater than 0.5: absorption atelectasis, O2 toxicity, or depression of ciliary or leukocyte function may occur
  • Increased FiO2 can worsen lung injury in patients with paraquat poisoning or patients receiving bleomycin
  • During laser bronchoscopy or tracheostomy: minimal FiO2 should be used to avoid intratracheal ignition
  • Fire hazard increased in the presence of high FiO2
  • Bacterial contamination can occur when nebulizers or humidifiers are used

Infant-Specific Hazards of Oxygen

  • Premature infants: PaO2 greater than 80 mmHg may contribute to Retinopathy of Prematurity (ROP). Retina do not vascularize completely, leading to detached retina and blindness
  • Certain congenital heart lesions such as hypoplastic left heart syndrome: high PaO2 can compromise the balance between pulmonary and systemic blood flow (duct dependent babies)
  • O2 flow directed at the face may stimulate an alteration in respiratory pattern

Low-Flow vs High-Flow Devices

CategoryFiO2Flow rates
Low-flow devicesVariable FiO2Generally less than 10 L/min
High-flow devicesFixed FiO2Generally greater than 10 L/min

Low-Flow Oxygen Devices

Flow is generally less than 10 L/min. FiO2 is variable and dependent upon the patient:

  • Tidal volume and respiratory rate (minute volume / min vol)
  • Increased Vt or RR decreases delivered FiO2 in low-flow devices
  • Inspiratory flow rate
  • Air entrainment around the device — increased air entrainment decreases FiO2

Flow Deficit Example

  • Patient inspiratory flow rate = 20 L/min
  • O2 device set at 8 L/min
  • Flow deficit = 12 L/min

The remaining 12 L/min comes from room air entrainment.

Low-Flow Device Summary

DeviceFlow rateFiO2
Nasal catheter1–5 L/min22%–35%
Nasal cannula0–6 L/minVariable (Rule of Fours)
Simple mask5–10 L/min (humidified up to 15 L/min)30%–60%
Non-rebreathing mask10 to greater than 15 L/min60%–80%
Trans-tracheal catheterLess than 6 L/min22%–35%

Nasal Catheter

  • Widely used as the standard low-flow oxygen delivery system until the late 1960s
  • Made of soft pliable PVC tube approximately 12 inches long with a series of small holes at the distal end
  • Adult catheter sizes: 12 and 14 Fr
  • Pediatric catheters: 8 and 10 Fr
  • Flow rates: 1 to 5 L/min
  • FiO2 ranges from 22% to 35%
  • Inserted blindly through the nose after length is estimated by measuring from the nose to the ear

Contraindications:

  • Nasal septum deviation
  • Mucosal congestion
  • Nasal polyps (abnormal growth of tissue projecting from a mucous membrane)

The nasal cannula and other less-invasive devices have replaced the nasal catheter.


Nasal Cannula

  • Most widely used device for administering low-flow oxygen to infants, children, and adults
  • Easily applied and well tolerated by most patients
  • Flow rates: 0 to 6 L/min
  • Must be humidified above 4 L/min

Rule of Fours

Each liter of flow equals approximately a 4% increase in FiO2:

20% + (O2 L/min x 4) = estimated FiO2
FlowEstimated FiO2
1 L/min24% (20 + 4)
2 L/min28% (20 + 8)
3 L/min32% (20 + 12)
4 L/min36% (20 + 16)
5 L/min40% (20 + 20)
6 L/min44% (20 + 24)

Simple Mask

Also called a face mask.

  • Covers the nose and mouth and is held in place by an elastic band around the head
  • Ambient air drawn in around the mask mixes with oxygen delivered to the mask, resulting in an approximate FiO2
  • Used when a nasal cannula is not appropriate, such as nasal obstruction
  • Flow rates: 5 to 10 L/min
  • FiO2: 30% to 60%
  • Humidified: up to 15 L/min

Safety Hazards (All Enclosed O2 Mask Systems)

  • CO2 accumulation and re-breathing of exhaled gases can occur at flow rates less than 5 L/min
  • All masks must be run at greater than 5 L/min to wash out exhaled CO2
  • Hypoxic rebound when the mask is removed
  • Asphyxiation secondary to aspiration
  • Patient compliance: difficulty eating, drinking, speaking on phone, and hygiene
  • Claustrophobia due to masks covering both the nose and mouth simultaneously

Non-Rebreathing Mask (NRB)

  • Incorporates reservoir bag and two one-way valves between the reservoir bag and mask
  • Prevents exhaled gas from entering the reservoir bag
  • Opens during inhalation to provide 100% oxygen
  • One exhalation port
  • During inhalation, the valve on the mask closes, reducing room air entrainment
  • Not humidified
  • FiO2: 60% to 80%
  • Flow rates: 10 to greater than 15 L/min
  • Must meet or exceed the patient inspiratory flow demand
  • Maintain inflated reservoir greater than 50%
  • Only device used to deliver Heliox

Trans-Tracheal Catheter

  • Oxygen delivered through a small catheter inserted transdermally
  • Application: long-term oxygen therapy
  • Flow rate less than 6 L/min
  • FiO2: 22% to 35%
  • Trachea acts as a reservoir to provide efficient use of O2 and reduced O2 flow
  • Hazards: site infections and mucus plugging

Small bore oxygen tubing connectors are used with this device.


High-Flow Oxygen Devices

Characteristics:

  • Flow rate exceeds the patient inspiratory flow rate
  • Delivers fixed FiO2
  • Increased flow rates minimize dilution with room air

Indications for high-flow oxygen:

  • Patient with high work of breathing
  • Accessory muscles in use
  • Increased tidal volume and/or respiratory rate

High-flow device types in this lesson:

  • Large volume nebulizer (LVN)
  • Air entrainment mask (Venturi mask)
  • HF nasal cannula (Vapotherm)

Large Volume Nebulizer (LVN)

  • FiO2: 30% to 100%
  • Jet nebulizer that also delivers aerosolized water
  • Uses large bore tubing to minimize flow resistance and prevent occlusion by condensate
  • Can be heated: hot plate, doughnut/yoke collar, or immersion heater

Delivery devices:

InterfaceClinical use
Aerosol maskGeneral aerosol delivery
Face tentDevice of choice post-extubation; alternative when a closed mask cannot be tolerated; FiO2 is less precise
Briggs T-pieceTracheostomy patients; weaning from mechanical ventilation
Tracheostomy collarTracheostomy delivery

For LVN use in bland aerosol and humidification therapy, see Humidification Delivery.


Venturi Mask (Air-Entrainment Mask)

Also called Venti mask.

  • Comprised of an aerosol mask with air entrainment adapters attached to small bore oxygen tubing
  • FiO2: 24% to 60%
  • Total flow can exceed 100 L/min due to air entrainment
  • Each adapter or dial is for a specific FiO2

Indications

  • Patients with increased minute volume requiring high flow to match or exceed inspiratory flow rate
  • If total flow to the mask is less than the patient inspiratory flow demand, delivered FiO2 will decrease
  • Occlusion of entrainment ports increases FiO2 while decreasing total flow — avoid obstructing air entrainment ports
  • Mouth breathing or blocked nasal passage
  • Tracheostomy patient transports through a trach collar
  • COPD patients who require FiO2 0.24 to 0.35 (24%–35%)

Total flow: combination of delivered oxygen flow rate plus entrained air flow rate, calculated for each FiO2.

Air-Oxygen Entrainment Ratios

FiO2Air : Oxygen
0.2425 : 1
0.2810 : 1
0.308 : 1
0.355 : 1
0.403 : 1
0.501.7 : 1
0.601 : 1

Remember: as air entrainment decreases, FiO2 increases. Do not obstruct air entrainment ports.

Calculating Air-Oxygen Entrainment Ratios

Air : O2 = (100 - % O2) : (% O2 - 21)

FiO2 = 24%:

Air: (100 - 24) / (24 - 21) = 76 / 3 = 25 : 1

FiO2 = 40%:

Air: (100 - 40) / (40 - 21) = 60 / 19 = 3 : 1

Calculation is possible only up to 60% maximum.


High-Flow Nasal Cannula (HFNC)

Vapotherm — high flow nasal cannula (HFNC):

  • Alternative to mask or CPAP oxygen therapy
  • Patient comfort increases compliance
  • Oxygen FiO2: 21% to 100%
  • Delivered flow rates:
    • Adults: 5–40 L/min
    • Neonates/children: 1–8 L/min
  • Heated (active) humidity incorporated with HFNC
  • Nasal delivery: prongs less than half inner diameter of nares

High velocity gas produces more pronounced clinical effects.

Exhalation effects:

  • CO2 washout
  • Dead space
  • Alveolar ventilation efficiency
  • Breathing efficiency

Positive pressure effects:

  • Mean airway pressure
  • Gas exchange
  • Work of breathing
  • Increased O2 concentration

Optimal humidification effects (compared to cool, dry gas):

  • Mucous clearance
  • Airway conductance
  • Pulmonary compliance

Oxygen Analyzers

Oxygen analyzers measure and monitor delivered FiO2.

Oxygen analyzers are necessary when specific FiO2 delivery must be known or is critical.

Examples: neonatal oxygen hoods, infant incubators, mechanical ventilators, high flow O2 devices, and others.

  • Perform 2-point calibration prior to each use
  • O2 error range is no greater than +/- 2% (if out of range, replace analyzer sensor or send for repair)

Types of Oxygen Analyzers

Common:

  • Electrochemical
  • Galvanic
  • Polarographic (Clark)
  • Paramagnetic
  • Zirconium cell
  • Raman scattering
  • Mass spectrometry

Specialized:

  • Galvanic fuel cell

Galvanic Fuel Cell

  • Cell runs continuously
  • Energy and accuracy depletes over time
  • Cap off sensor when not in use
  • Slow response time
  • Accuracy +/- 2% or replace cell

Operation:

  • Chemical reaction creates electrical current
  • Gold cathode (+), lead anode (-) immersed in electrolyte gel
  • PiO2 measured then converted to read as FiO2

Polarographic (Clark) Electrode

  • Operates continuously or intermittently
  • Fast response time
  • Accuracy +/- 2% or replace cell

Operation:

  • Battery powered source (9 Volt)
  • Platinum cathode (+), silver anode (-)
  • PiO2 measured then converted to read as FiO2

Electrochemical O2 Analyzer Comparison

FeatureGalvanic fuel cellPolarographic (Clark)
ResponseSlow (< 60 sec)Faster (< 30 sec)
Anesthetic gasesSafeSafe
PowerElectrochemical (no batteries)Battery replacement
OperationContinuous; consumes O2Intermittent
FiO2 falsely increased byTemp increase, pressure increaseTemp increase, pressure increase
FiO2 falsely decreased byHumidity decrease, altitude decreaseHumidity decrease, altitude decrease

Analyzer Calibration Procedure

  1. Assemble: Connect cable to O2 cell, then attach to analyzer unit. Turn on unit to verify lights and audible alarms.
  2. Two-point calibration (21% and 100%):
    • Calibrate to room air — press "21%" then "unlock"
    • Calibrate to 100% — place sensor in plastic bag and bleed oxygen into bag. Press "100%" then "unlock"
  3. Set high/low alarms: Press "set", then "up arrow" for high alarm. Press "set", then "down arrow" for low alarm.
  4. Place analyzer in line with gas source with Briggs adaptor.
  5. Ventilator use:
    • Place adaptor pre-heater (cool)/humidifier (dry)
    • Recalibrate with altitude changes
    • Set O2 alarm (if available/default): high 10% above FiO2, low 10% below FiO2

High-Yield Review

Indications for Oxygen Therapy

  • Documented hypoxemia confirmed by ABG: PaO2 less than 60 mmHg or SaO2 less than 90% on room air (21% FiO2); or below desirable range for specific clinical situations
  • Acute care when hypoxemia is suspected: cor pulmonale, congestive heart failure, severe trauma, acute MI
  • Short-term therapy or post-anesthesia recovery

Low-Flow Devices

  • Variable approximate FiO2; flow generally less than 10 L/min
  • FiO2 depends on minute ventilation (Ve = respiratory rate x tidal volume)
  • Increasing minute ventilation increases air entrainment and decreases FiO2

High-Flow Devices

  • Fixed known FiO2; flow generally greater than 10 L/min
  • Total flow exceeds patient inspiratory demand
  • Increasing flow minimizes room air dilution

Device Flow and FiO2 Ranges

DeviceFlowFiO2
Nasal catheter1–5 L/min22%–35%
Nasal cannula0–6 L/minRule of Fours estimate
Simple mask5–10 L/min30%–60%
Non-rebreathing mask10 to greater than 15 L/min60%–80%
Trans-tracheal catheterless than 6 L/min22%–35%
Venturi maskHigh total flow24%–60% fixed
HFNC (Vapotherm)Adults 5–40 L/min; peds/neonates 1–8 L/min21%–100%

Oxygen Analyzer Calibration

  • 2-point calibration at 21% and 100% before each use
  • Accuracy within +/- 2%