Oxygen Delivery Devices
Objectives - Describe indications for oxygen therapy; administer oxygen utilizing low-flow and high-flow therapy devices; describe flow rate and FiO2 range for each oxygen device; perform a two-point calibration on an oxygen analyzer.
Indications for Oxygen
Documented Hypoxemia
Confirmed by ABG:
- PaO2 less than 60 mmHg or SaO2 less than 90% on room air (21% FiO2)
- PaO2 or SaO2 below desirable range for specific clinical situations, such as pneumonia
Acute Care Situations in Which Hypoxemia Is Suspected
Examples include:
- Cor pulmonale
- Congestive heart failure
- Severe trauma (severe blood loss)
- Acute myocardial infarction
Short-Term Therapy or Surgical Intervention
- Post-anesthesia recovery
Contraindications for Oxygen
With few exceptions, no specific contraindication to O2 therapy exists when indications are present.
Certain delivery devices are contraindicated:
- Nasal cannulas and nasopharyngeal catheters in pediatric and neonatal patients with obstruction
Hazards of Oxygen
- PaO2 greater than or equal to 60 mmHg: ventilatory depression may occur rarely in spontaneously breathing patients with elevated PaCO2 (COPD)
- With FiO2 greater than 0.5: absorption atelectasis, O2 toxicity, or depression of ciliary or leukocyte function may occur
- Increased FiO2 can worsen lung injury in patients with paraquat poisoning or patients receiving bleomycin
- During laser bronchoscopy or tracheostomy: minimal FiO2 should be used to avoid intratracheal ignition
- Fire hazard increased in the presence of high FiO2
- Bacterial contamination can occur when nebulizers or humidifiers are used
Infant-Specific Hazards of Oxygen
- Premature infants: PaO2 greater than 80 mmHg may contribute to Retinopathy of Prematurity (ROP). Retina do not vascularize completely, leading to detached retina and blindness
- Certain congenital heart lesions such as hypoplastic left heart syndrome: high PaO2 can compromise the balance between pulmonary and systemic blood flow (duct dependent babies)
- O2 flow directed at the face may stimulate an alteration in respiratory pattern
Low-Flow vs High-Flow Devices
| Category | FiO2 | Flow rates |
|---|---|---|
| Low-flow devices | Variable FiO2 | Generally less than 10 L/min |
| High-flow devices | Fixed FiO2 | Generally greater than 10 L/min |
Low-Flow Oxygen Devices
Flow is generally less than 10 L/min. FiO2 is variable and dependent upon the patient:
- Tidal volume and respiratory rate (minute volume / min vol)
- Increased Vt or RR decreases delivered FiO2 in low-flow devices
- Inspiratory flow rate
- Air entrainment around the device — increased air entrainment decreases FiO2
Flow Deficit Example
- Patient inspiratory flow rate = 20 L/min
- O2 device set at 8 L/min
- Flow deficit = 12 L/min
The remaining 12 L/min comes from room air entrainment.
Low-Flow Device Summary
| Device | Flow rate | FiO2 |
|---|---|---|
| Nasal catheter | 1–5 L/min | 22%–35% |
| Nasal cannula | 0–6 L/min | Variable (Rule of Fours) |
| Simple mask | 5–10 L/min (humidified up to 15 L/min) | 30%–60% |
| Non-rebreathing mask | 10 to greater than 15 L/min | 60%–80% |
| Trans-tracheal catheter | Less than 6 L/min | 22%–35% |
Nasal Catheter
- Widely used as the standard low-flow oxygen delivery system until the late 1960s
- Made of soft pliable PVC tube approximately 12 inches long with a series of small holes at the distal end
- Adult catheter sizes: 12 and 14 Fr
- Pediatric catheters: 8 and 10 Fr
- Flow rates: 1 to 5 L/min
- FiO2 ranges from 22% to 35%
- Inserted blindly through the nose after length is estimated by measuring from the nose to the ear
Contraindications:
- Nasal septum deviation
- Mucosal congestion
- Nasal polyps (abnormal growth of tissue projecting from a mucous membrane)
The nasal cannula and other less-invasive devices have replaced the nasal catheter.
Nasal Cannula
- Most widely used device for administering low-flow oxygen to infants, children, and adults
- Easily applied and well tolerated by most patients
- Flow rates: 0 to 6 L/min
- Must be humidified above 4 L/min
Rule of Fours
Each liter of flow equals approximately a 4% increase in FiO2:
20% + (O2 L/min x 4) = estimated FiO2
| Flow | Estimated FiO2 |
|---|---|
| 1 L/min | 24% (20 + 4) |
| 2 L/min | 28% (20 + 8) |
| 3 L/min | 32% (20 + 12) |
| 4 L/min | 36% (20 + 16) |
| 5 L/min | 40% (20 + 20) |
| 6 L/min | 44% (20 + 24) |
Simple Mask
Also called a face mask.
- Covers the nose and mouth and is held in place by an elastic band around the head
- Ambient air drawn in around the mask mixes with oxygen delivered to the mask, resulting in an approximate FiO2
- Used when a nasal cannula is not appropriate, such as nasal obstruction
- Flow rates: 5 to 10 L/min
- FiO2: 30% to 60%
- Humidified: up to 15 L/min
Safety Hazards (All Enclosed O2 Mask Systems)
- CO2 accumulation and re-breathing of exhaled gases can occur at flow rates less than 5 L/min
- All masks must be run at greater than 5 L/min to wash out exhaled CO2
- Hypoxic rebound when the mask is removed
- Asphyxiation secondary to aspiration
- Patient compliance: difficulty eating, drinking, speaking on phone, and hygiene
- Claustrophobia due to masks covering both the nose and mouth simultaneously
Non-Rebreathing Mask (NRB)
- Incorporates reservoir bag and two one-way valves between the reservoir bag and mask
- Prevents exhaled gas from entering the reservoir bag
- Opens during inhalation to provide 100% oxygen
- One exhalation port
- During inhalation, the valve on the mask closes, reducing room air entrainment
- Not humidified
- FiO2: 60% to 80%
- Flow rates: 10 to greater than 15 L/min
- Must meet or exceed the patient inspiratory flow demand
- Maintain inflated reservoir greater than 50%
- Only device used to deliver Heliox
Trans-Tracheal Catheter
- Oxygen delivered through a small catheter inserted transdermally
- Application: long-term oxygen therapy
- Flow rate less than 6 L/min
- FiO2: 22% to 35%
- Trachea acts as a reservoir to provide efficient use of O2 and reduced O2 flow
- Hazards: site infections and mucus plugging
Small bore oxygen tubing connectors are used with this device.
High-Flow Oxygen Devices
Characteristics:
- Flow rate exceeds the patient inspiratory flow rate
- Delivers fixed FiO2
- Increased flow rates minimize dilution with room air
Indications for high-flow oxygen:
- Patient with high work of breathing
- Accessory muscles in use
- Increased tidal volume and/or respiratory rate
High-flow device types in this lesson:
- Large volume nebulizer (LVN)
- Air entrainment mask (Venturi mask)
- HF nasal cannula (Vapotherm)
Large Volume Nebulizer (LVN)
- FiO2: 30% to 100%
- Jet nebulizer that also delivers aerosolized water
- Uses large bore tubing to minimize flow resistance and prevent occlusion by condensate
- Can be heated: hot plate, doughnut/yoke collar, or immersion heater
Delivery devices:
| Interface | Clinical use |
|---|---|
| Aerosol mask | General aerosol delivery |
| Face tent | Device of choice post-extubation; alternative when a closed mask cannot be tolerated; FiO2 is less precise |
| Briggs T-piece | Tracheostomy patients; weaning from mechanical ventilation |
| Tracheostomy collar | Tracheostomy delivery |
For LVN use in bland aerosol and humidification therapy, see Humidification Delivery.
Venturi Mask (Air-Entrainment Mask)
Also called Venti mask.
- Comprised of an aerosol mask with air entrainment adapters attached to small bore oxygen tubing
- FiO2: 24% to 60%
- Total flow can exceed 100 L/min due to air entrainment
- Each adapter or dial is for a specific FiO2
Indications
- Patients with increased minute volume requiring high flow to match or exceed inspiratory flow rate
- If total flow to the mask is less than the patient inspiratory flow demand, delivered FiO2 will decrease
- Occlusion of entrainment ports increases FiO2 while decreasing total flow — avoid obstructing air entrainment ports
- Mouth breathing or blocked nasal passage
- Tracheostomy patient transports through a trach collar
- COPD patients who require FiO2 0.24 to 0.35 (24%–35%)
Total flow: combination of delivered oxygen flow rate plus entrained air flow rate, calculated for each FiO2.
Air-Oxygen Entrainment Ratios
| FiO2 | Air : Oxygen |
|---|---|
| 0.24 | 25 : 1 |
| 0.28 | 10 : 1 |
| 0.30 | 8 : 1 |
| 0.35 | 5 : 1 |
| 0.40 | 3 : 1 |
| 0.50 | 1.7 : 1 |
| 0.60 | 1 : 1 |
Remember: as air entrainment decreases, FiO2 increases. Do not obstruct air entrainment ports.
Calculating Air-Oxygen Entrainment Ratios
Air : O2 = (100 - % O2) : (% O2 - 21)
FiO2 = 24%:
Air: (100 - 24) / (24 - 21) = 76 / 3 = 25 : 1
FiO2 = 40%:
Air: (100 - 40) / (40 - 21) = 60 / 19 = 3 : 1
Calculation is possible only up to 60% maximum.
High-Flow Nasal Cannula (HFNC)
Vapotherm — high flow nasal cannula (HFNC):
- Alternative to mask or CPAP oxygen therapy
- Patient comfort increases compliance
- Oxygen FiO2: 21% to 100%
- Delivered flow rates:
- Adults: 5–40 L/min
- Neonates/children: 1–8 L/min
- Heated (active) humidity incorporated with HFNC
- Nasal delivery: prongs less than half inner diameter of nares
High velocity gas produces more pronounced clinical effects.
Exhalation effects:
- CO2 washout
- Dead space
- Alveolar ventilation efficiency
- Breathing efficiency
Positive pressure effects:
- Mean airway pressure
- Gas exchange
- Work of breathing
- Increased O2 concentration
Optimal humidification effects (compared to cool, dry gas):
- Mucous clearance
- Airway conductance
- Pulmonary compliance
Oxygen Analyzers
Oxygen analyzers measure and monitor delivered FiO2.
Oxygen analyzers are necessary when specific FiO2 delivery must be known or is critical.
Examples: neonatal oxygen hoods, infant incubators, mechanical ventilators, high flow O2 devices, and others.
- Perform 2-point calibration prior to each use
- O2 error range is no greater than +/- 2% (if out of range, replace analyzer sensor or send for repair)
Types of Oxygen Analyzers
Common:
- Electrochemical
- Galvanic
- Polarographic (Clark)
- Paramagnetic
- Zirconium cell
- Raman scattering
- Mass spectrometry
Specialized:
- Galvanic fuel cell
Galvanic Fuel Cell
- Cell runs continuously
- Energy and accuracy depletes over time
- Cap off sensor when not in use
- Slow response time
- Accuracy +/- 2% or replace cell
Operation:
- Chemical reaction creates electrical current
- Gold cathode (+), lead anode (-) immersed in electrolyte gel
- PiO2 measured then converted to read as FiO2
Polarographic (Clark) Electrode
- Operates continuously or intermittently
- Fast response time
- Accuracy +/- 2% or replace cell
Operation:
- Battery powered source (9 Volt)
- Platinum cathode (+), silver anode (-)
- PiO2 measured then converted to read as FiO2
Electrochemical O2 Analyzer Comparison
| Feature | Galvanic fuel cell | Polarographic (Clark) |
|---|---|---|
| Response | Slow (< 60 sec) | Faster (< 30 sec) |
| Anesthetic gases | Safe | Safe |
| Power | Electrochemical (no batteries) | Battery replacement |
| Operation | Continuous; consumes O2 | Intermittent |
| FiO2 falsely increased by | Temp increase, pressure increase | Temp increase, pressure increase |
| FiO2 falsely decreased by | Humidity decrease, altitude decrease | Humidity decrease, altitude decrease |
Analyzer Calibration Procedure
- Assemble: Connect cable to O2 cell, then attach to analyzer unit. Turn on unit to verify lights and audible alarms.
- Two-point calibration (21% and 100%):
- Calibrate to room air — press "21%" then "unlock"
- Calibrate to 100% — place sensor in plastic bag and bleed oxygen into bag. Press "100%" then "unlock"
- Set high/low alarms: Press "set", then "up arrow" for high alarm. Press "set", then "down arrow" for low alarm.
- Place analyzer in line with gas source with Briggs adaptor.
- Ventilator use:
- Place adaptor pre-heater (cool)/humidifier (dry)
- Recalibrate with altitude changes
- Set O2 alarm (if available/default): high 10% above FiO2, low 10% below FiO2
High-Yield Review
Indications for Oxygen Therapy
- Documented hypoxemia confirmed by ABG: PaO2 less than 60 mmHg or SaO2 less than 90% on room air (21% FiO2); or below desirable range for specific clinical situations
- Acute care when hypoxemia is suspected: cor pulmonale, congestive heart failure, severe trauma, acute MI
- Short-term therapy or post-anesthesia recovery
Low-Flow Devices
- Variable approximate FiO2; flow generally less than 10 L/min
- FiO2 depends on minute ventilation (Ve = respiratory rate x tidal volume)
- Increasing minute ventilation increases air entrainment and decreases FiO2
High-Flow Devices
- Fixed known FiO2; flow generally greater than 10 L/min
- Total flow exceeds patient inspiratory demand
- Increasing flow minimizes room air dilution
Device Flow and FiO2 Ranges
| Device | Flow | FiO2 |
|---|---|---|
| Nasal catheter | 1–5 L/min | 22%–35% |
| Nasal cannula | 0–6 L/min | Rule of Fours estimate |
| Simple mask | 5–10 L/min | 30%–60% |
| Non-rebreathing mask | 10 to greater than 15 L/min | 60%–80% |
| Trans-tracheal catheter | less than 6 L/min | 22%–35% |
| Venturi mask | High total flow | 24%–60% fixed |
| HFNC (Vapotherm) | Adults 5–40 L/min; peds/neonates 1–8 L/min | 21%–100% |
Oxygen Analyzer Calibration
- 2-point calibration at 21% and 100% before each use
- Accuracy within +/- 2%