Oxygen & Mixed Gas Therapy

Therapeutic respiratory gases, medical gas storage and delivery, cylinder safety systems, regulators, flowmeters, air-oxygen blending, cylinder duration, Heliox delivery, and oxygen concentrators.

Cylinder Duration Calculation Practice

Calculate each duration in minutes. Round to the nearest whole minute.

Formula

[(PSI - 500) x tank factor] / L/min = minutes

The 500 psig safety factor leaves reserve time before the cylinder runs empty.

Heliox H: (PSI x 2.50) / L/min = minutes

Tank Factors

D

0.16 L/psig

E

0.28 L/psig

G

2.41 L/psig

H & K

3.14 L/psig

Heliox H

2.50 L/psig

Exercise 1

A patient is receiving oxygen by nasal cannula during transport.

Tank

E

Gas

Oxygen

PSI

2000

Flow

5

Exercise 2

A small cylinder is being used for low-flow oxygen delivery.

Tank

D

Gas

Oxygen

PSI

1800

Flow

2

Exercise 3

A large cylinder is powering a high-flow oxygen device.

Tank

H/K

Gas

Oxygen

PSI

2200

Flow

10

Exercise 4

A G cylinder is being used for a transport device at 15 L/min.

Tank

G

Gas

Oxygen

PSI

1600

Flow

15

Exercise 5

An E cylinder has reached 1000 psig and still needs reserve time.

Tank

E

Gas

Oxygen

PSI

1000

Flow

4

Exercise 6

An H cylinder of Heliox is running through a non-rebreather mask.

Tank

Heliox H

Gas

Heliox

PSI

2200

Flow

10

Exercise 7

An H cylinder of Heliox is being used with a ventilator setup.

Tank

Heliox H

Gas

Heliox

PSI

1800

Flow

12

Current score: 0 / 7

Answers within 1 minute of the rounded result count as correct.

Heliox Delivery with O2 Flowmeters

When Heliox is delivered with an oxygen flowmeter, actual flow is greater than the indicated flow.

80/20

O2 flow x 1.8

70/30

O2 flow x 1.6

60/40

O2 flow x 1.4

Flow Exercise 1

Heliox 80/20 is set on an oxygen flowmeter at 10 L/min. Calculate the actual delivered flow.

Mix

80/20

O2 Flow

10

Factor

1.8

Flow Exercise 2

Heliox 70/30 is set on an oxygen flowmeter at 8 L/min. Calculate the actual delivered flow.

Mix

70/30

O2 Flow

8

Factor

1.6

Flow Exercise 3

Heliox 60/40 is set on an oxygen flowmeter at 12 L/min. Calculate the actual delivered flow.

Mix

60/40

O2 Flow

12

Factor

1.4

Heliox flow score: 0 / 3

Answers within 0.5 L/min of the calculated flow count as correct.